Charging the Patient Simulator
- Battery charging time is approximately 3 hours. Once fully charged, the batteries can last for up to 4 hours. 
- Plug the Patient Simulator to the external power supply. Turn the Patient Simulator on to charge the batteries. 
- Battery and power status can be viewed on the Patient Simulator’s power panel and the Simulator Status window in LLEAP. 
CAUTIONS
- Do not run the Patient Simulator for more than 1 minute on a single battery. 
- After the Patient Simulator is turned off, wait 20 seconds before restarting or the Patient Simulator may not function properly. 
- Never store empty batteries. 
- Only use the SimMan Family (except SimMan ALS) external power supply and batteries. 
- The external power supply is for indoor use only. 
- Do not store fully charged batteries for longer than a month. 
NOTES
- The Patient Simulator will automatically shut down if battery temperature rises above 60 °C (140 °F) or the remaining charge falls below 6% on one of the two batteries. 
- If both batteries are removed while the Patient Simulator is running on batteries, the Patient Simulator will shut down and simulation data will be lost. 
- Refer to the Important Product Information for further information.