Head, Neck and Airway
Replacing cricothyroid tape and neck skin
After creating an emergency airway through the cricothyroid membrane, replace the perforated tape before starting a new simulation session.
- Remove the neck skin. 
- Remove damaged or perforated tape. 
- Select a new section of cricothyroid tape. 
- Replace with new tape, ensuring it completely covers and seals the opening to prevent leakage during ventilation. 
- Replace the neck skin. 
NOTE
If the used neck skin is in good condition, move the skin along to position a new section over the cricothyroid tape.